CS QUESTIONS • CS ANSWERS


IUSS cycle frequency; dif ferentiating sanitization, disinfection and sterilization


by Ray Taurasi Q


are acceptable to run during any specifi c time- frame. Sterilization process failures are a ma- jor concern and pose a serious risk to patient safety and well-being. It makes more sense to place great emphasis on each and every sterilization cycle regardless of whether the cycle is IUSS or a traditional terminal cycle. It is imperative that every single step and detail in the reprocessing/sterilization pro- cess is performed precisely in accordance with the medical device and equipment manufac- turers’ instruction for use (IFU). There are absolutely no short cuts in the sterilization process. All policies and procedures must be followed for cleaning, disinfection, inspection, assembly, instrument positioning, packaging, loading sterilizer, sterilizer operation, cycle settings, monitoring, documentation, and adherence to aseptic techniques. Clearly, sterilization is a complex and very precise process and one that cannot be rushed. For that reason all professional entities agree and recommend that the use of IUSS be reserved for true emergency and/ or unanticipated events. Having inadequate instrumentation to meet surgical case load needs is not considered an emergency or un- anticipated event. Surveyors from CMS, TJC, and other accrediting bodies will be looking at the effi cacy of the entire sterilization pro- cess, not merely how many IUSS cycles are run. They will also assess your adherence to professional guidelines and standards relative to managing and monitoring all sterilization processing, including IUSS. As a quality improvement initiative, you may want to consider benchmarking your own IUSS cycle rates against themselves to demonstrate that you are managing the pro- cess and implementing measures to reserve the use of IUSS to the unanticipated events.


A


How many Immediate Use Steam Ster- ilization (IUSS) cycles are acceptable during a one-month timeframe? I have not been able to fi nd that information in any of the AAMI or AORN guidelines. In survey- ing my colleagues at various hospitals I have gotten a multitude of different answers. We have an accreditation inspection coming up and I don’t want to be cited for not following regulations.


There are no regulations, guidelines or standards stating how many IUSS cycles


SUBMIT YOUR QUESTIONS email: editor@hpnonline.com phone: (941) 927-9345 ext. 202


Calculate your IUSS rates by dividing the number of IUSS cycles per month by the number of procedures per month.


Q A


Is it essential to sanitize all soiled items returned to CS, or is decontamination/ disinfection suffi cient prior to sterilization?


It is important to realize that decontami- nation (cleaning), sanitization, disinfec- tion and sterilization are each different and unique processes. The fi rst step in reprocess- ing involves thorough cleaning which is the removal of soil from used materials. The sec- ond step involves a microbicidal process such as sanitization, or disinfection. The objective of decontamination is to protect individuals from disease caused by contact with patho- genic organisms on soiled medical devices. Thorough cleaning, while an essential fi rst step, may not always be suffi cient to ensure that an item is safe to handle. Currently, there is no standard to measure just how “clean” or safe-to-handle an object must be. There- fore, to ensure that harmful microorganisms are destroyed, a microbicidal process may be employed. Medical devices that have been contaminated by blood, body fluids or large microbial populations must always be subjected to a microbicidal process after cleaning. Deciding which process to use is a risk-versus-benefi t decision for each item you process. Therefore, a clear understanding of these processes is imperative.


Sanitization is suitable for easily-killed microorganisms and can reduce the number of microbial contaminants on an inanimate surface to a relatively safe level. Following thorough cleaning, the objects are rinsed in hot water or steam-purged for a designated period of time depending on method used. Examples of sanitizing equipment include cart washes, steam guns, and dishwashers. Sanitization is adequate for items that only come in contact with the surface of unbroken skin. Disinfection provides a higher level of safety and can be used on work surfaces, medical devices and equipment that have come in contact with highly contaminated substances, body fl uids and blood. The disin-


30 January 2016 • HEALTHCARE PURCHASING NEWS • hpnonline.com


fection process may be accomplished either by a thermal or chemical exposure. Pasteuriza- tion is an example of a thermal disinfection process. It involves exposing an object to a hot water bath at between 150 and 170 degrees Fahrenheit for 30 minutes. Contact with wa- ter at or above 180 to 205 degrees Fahrenheit for one minute can provide an intermediate level of disinfection. This is the process used in automated instrument washers during the fi nal rinse.


Chemical disinfection processes employ various liquid chemicals that contain agents such as quaternary ammonium compounds, iodophors, hydrogen peroxide, phenolics, chlorine compounds and glutaraldehydes. Chemical disinfection can be a complicated process, and careful selection of the appropri- ate chemical for each medical device or piece of equipment is imperative. Labeling will identify a chemical’s active ingredients and concentrations and provide the scope and spectrum of microbicidal activity. Directions for use must be followed precisely, including required contact time and temperature. It is essential to follow the device manufacturer’s IFU for proper cleaning and disinfection. As noted previously, thorough cleaning is a critical step and if not done effectively, can impede the effi cacy of the disinfection pro- cess. To monitor the effi cacy of the cleaning process all automated washers (e.g., instru- ment washers, cart washers, ultra-sonics and automated endoscope reprocessor) should be tested weekly, preferably daily, in accordance with AAMI ST79 standards. There are specifi c testing tools and devices available for the various automated washers which verify the performance of the equipment. The cleaning effi cacy of medical devices can be verifi ed with testing devices that detect residuals of organic matter such as, blood, carbohydrates, and proteins. HPN


Ray Taurasi is Eastern Regional Director of Clini- cal Sales and Services for Healthmark Industries. His healthcare career spans over three decades as an Administrator, Educator, Technologist and Consultant. He is a member of AORN, AHA, SGNA, AAMI and a past president of IAHCSMM. Taurasi has been a faculty member of numerous colleges teaching in the divisions of business administration and health sciences.


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46