IAHCSMM VIEWPOINT


Beware these 5 common instrument use, inspection and testing myths


by Rick Schultz M


isinformation about surgical instruments and instrument pro- cessing is widespread across


the healthcare industry. Many educators have not updated their knowledge and teaching techniques to stay current with the instrument manufacturer’s knowledge and testing methods. As a result, outdated processing knowledge and inaccurate edu- cation may result in patient safety issues. Knowing the facts is essential.


What follows are some common myths that must be addressed:


Myth #1: It is safe and acceptable to use hemostats as tube clamps for intravenous tubing. Fact: No. Using a hemostat to clamp the tube is not safe for the patient. Hemostats typically have fi ner aws than a tube clamp, and the box lock of hemostats are not as strong. Because of this, hemostats are highly prone to cracking and, even worse, releas- ing/popping off the tube. Hemostats are designed to clamp vessels to control blood fl ow, which is achieved by both aws clos- ing fully. Furthermore, the serrations on a hemostat may damage the actual tubing. Common tube clamps, such as Presbyte- rian, come with smooth or serrated aws to secure tubing. The other most popular tube clamp is the Vorse, which has a “guard” to support the aws and box loc when clamp- ing the tube. This pattern is called a Vorse with guard or Vorse Guardian.


Myth #2: To inspect a needle holder, it is necessary to click it on the fi rst ratchet and hold it up to the light to visualize the center of the jaws. Fact: This technique is outdated, slow and reduces effi ciency when inspecting surgical instruments. Because there are different teeth per square inch, depending on the pattern, the needle holder may be worn, even though light does not show through; however, this technique is effective for needle holders with smooth aws, such as Webster, alsey and Castrovieo, which have no tread in the aws. he most recog- nized and effi cient technique for inspecting


needle holder aws is to separate the rings and visually inspect the hinge area and the aws for treadwear and bioburden.


Myth #3: Clamping aluminum foil is the preferred way to test needle holders for jaw wear. Fact: Using aluminum foil to test tread wear is a slow, outdated technique. The most effective method is to separate the rings and inspect the aws for treadwear and bioburden.


Myth #4: Ratchet testing is performed by tapping the instrument in the palm of the hand. Fact: Tapping the instrument in the palm of the hand is not an accurate method to ensure the ratchets are properly aligned. The hand has a soft surface and is not completely fl at. A completely fl at surface must be used, such as tapping on a table/ work surface. The purpose of tapping the instrument on a hard work surface is to simulate the instrument hitting another instrument, causing it to pop open. When performing this test, click the ratchet only once so it is in the fi rst position. f the ratchet pops open, that indicates the ratchets need to be realigned, and the instrument should be removed from service and sent out for repair. This is a common and inexpensive repair. This technique should be used on all hemostats, needle holders and any other ratcheted instrument.


of the aw. he instrument should not tear or snag the testing card. The middle and bac of the cup aw are not used during the surgical procedure. When testing the sharpness of a scissor, either red or yellow scissor test material should be used. Scis- sors with an overall length longer than 4 ½” should cut cleanly through red scissor test material. Scissors with an overall length of 4 ½” and shorter should cut cleanly through yellow scissor test material. When testing the sharpness of a scissor, use only the distal 1/3 of the blade (it is incorrect to test the middle or bac of the blade). f the testing material snags or is not cut cleanly, the scissors should be tagged and sent out for sharpening.


Conclusion


Although many of these myths, initially, may seem small and insignifi cant, they can have a maor impact on the performance of the instrument and, more importantly, on the patient. Knowing the facts will improve patient safety. HPN


Myth #5: When testing rongeurs and scissors for sharpness, the entire jaw or front half of the jaw or blade should be tested. Fact: Rongeurs should cut cleanly through one thickness of an index card or card stock. To perform this test, use only the top 1/3


30 May 2020 • HEALTHCARE PURCHASING NEWS • hpnonline.com


Rick Schultz is an author, inventor and lecturer, and the retired CEO of Spectrum Surgical Instruments Corp.


An extended version of this article was published in the Jan/Feb 2019 issue of PROCESS, IAHC- SMM’s bi-monthly magazine.


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58